Indication error is signed deviation
Compare the reading with the reference angle from a polygon, an interferometric angle base or a calibrated table; the difference is the indication error and must keep its sign.
Only the shape of that error curve tells you whether a specific working point is usable. A single ±1″ collapses the whole curve into one number.
Linearity is the shape of that curve
Uniform sampling across the range shows whether the error grows proportionally (scale or focal length) or is worst at the ends (uncorrected sine error).
Always state range and number of points, or vendor figures cannot be compared.
Repeatability tells you whether one reading is enough
Read n times at the same position without re-clamping and report the range or standard deviation. Poor repeatability makes linearity irrelevant; check source, airflow, bracket and stability criterion first.
A working rule: repeatability no worse than one fifth of the tolerance, otherwise measurement dominates the uncertainty.