Home / Polygons: dividing the circle into a traceable angle standard
PRINCIPLE NOTE

Polygons: dividing the circle into a traceable angle standard

A polygon exploits circle closure: the n face-normal angles must sum to 360°, so the polygon calibrates itself without a higher-order angle standard. It is the key artefact for tracing rotary tables and goniometers.

DUAL AXIS+Y-Y+X field-X field
Σθᵢ = 360° (closure); nominal adjacent-face angle = 360°/n

Animated explanation

AutocollimatorNominal facet angles ± deviationFull-circle closure: Σ deviations → 0Schematic · not to scale · not measured data
1 / 4
  1. Establish the geometry
  2. See how one reading forms
  3. Change a condition and observe
  4. Return to the formula and its limits

The animation explains relationships and directions. Read numbers from the formula, the standard and your records.

Closure: why a polygon calibrates itself

After reading every face with an autocollimator and a rotary table, distributing the closure error over the faces separates the table systematic error from the polygon deviation itself. That is the full-closure method; no face truth value is needed first.

More faces mean denser index points per revolution; 12, 18, 23 and 36 faces are common. Grade-0 polygons hold adjacent-face deviation within about ±1″.

In practice: polygon, spindle and autocollimator setup

The polygon mounts on the table or goniometer spindle; eccentricity introduces a second-harmonic error, so centre the arbor to a few micrometres. The autocollimator stands on an independent support, axis aimed at face-centre height, and does not rotate.

Read after each face settles, repeat the revolution 2–3 times and average. Temperature drift slowly grows the closure error, so keep each revolution short.

Using the result: error table and compensation

The output is an indexing error table: the deviation at each angle point. CNC rotary tables can load it for error compensation; goniometers use it to correct factory scales; laboratories use it to calibrate lower-grade artefacts.

Boundary: the polygon method gives closure-referenced indexing error and does not replace spindle runout tests (radial/axial).