Closure: why a polygon calibrates itself
After reading every face with an autocollimator and a rotary table, distributing the closure error over the faces separates the table systematic error from the polygon deviation itself. That is the full-closure method; no face truth value is needed first.
More faces mean denser index points per revolution; 12, 18, 23 and 36 faces are common. Grade-0 polygons hold adjacent-face deviation within about ±1″.
In practice: polygon, spindle and autocollimator setup
The polygon mounts on the table or goniometer spindle; eccentricity introduces a second-harmonic error, so centre the arbor to a few micrometres. The autocollimator stands on an independent support, axis aimed at face-centre height, and does not rotate.
Read after each face settles, repeat the revolution 2–3 times and average. Temperature drift slowly grows the closure error, so keep each revolution short.
Using the result: error table and compensation
The output is an indexing error table: the deviation at each angle point. CNC rotary tables can load it for error compensation; goniometers use it to correct factory scales; laboratories use it to calibrate lower-grade artefacts.
Boundary: the polygon method gives closure-referenced indexing error and does not replace spindle runout tests (radial/axial).